    KLEINWACHTER CPM374 CHARGED PLATE MONITOR 
    with integrated Electro-Fieldmeter 
    SYSTEM PROF. DR. ING. HANS KLEINW?CHTER 
     
    The CHARGE PLATE Monitor 374 measures the voltage potential on its integrated Plate-Electrode according the Fieldmill-Influence-Principal and is suitable to determine the Static Decay Time from a selected higher voltage threshold to a lower voltage threshold level. 
      
    CPM374 
    Forschungs-. Entwlcklungs-. Produktlons-u. Vertrlebsges.m.b.H
     
    The CHARGE PLATE Monitor CPM374 consists of an integrated timer, a high-voltage unit and an electro-fieldmill-sensor and achieves therefore the high impedance of 10
    15 W to measure electrostatic voltage potentials and static decay times of air ionizer equipment. 
     
    The unit meets the requirements of the Standard IEC 61340-5-1 and the American Standard EOS/ESD S.3.1.  
     
    For the vertical measurement use the included three Distance pieces. Screw them on the backside of the unit. 
     
    Important: The high voltage unit is protected by a high ohmic current limiting resistor and therefore there is no danger when touching the charged plate. 
     
    The CPM374 is battery powered and due to its compact design, it is very suitable for the use in industrial areas. It can also be operated at a power supply (230/115 VAC). The CPM374 is menu-driven and will be controlled by means of 2 switches for simple operation. All selected parameters will be shown in the 2 rows LCD-display for a better orientation of the operator.  
     
    The enclosed software enables the remote control operation of the CPM374 and data processing by means of a PC. 
     
    2.1  Switching on  
    The CHARGE PLATE Monitor CPM374 is equipped with a NiCd-battery and ready for operation - use the enclosed power supply only to charge the battery. The display background illumination is not operating when battery powered. The battery operation features an automatic switch-off function, if the switches were not touched for 
    approximay 10 minutes. when the unit is run on the power supply (230/115 VAC) the background illumination is switched on while the automatic switch-off function is disabled. 
     
    Press button B to switch on the CPM374. 
      
    1.  Description 
    2.  Operation 
    2.2  Decay Time and Offset Voltage 
     
    The software version will be shown shortly before the measurement mode and the command to initiate the measurement appear on the display: 
     
    DECAY TIME: AUTO 
    PRESS START! 
     
    The initially setting is saved in the EEPROM and will be shown after each switch-on. 
     
    If the auto-mode is selected, as indicated in the example above, then the time will be measured for the positive voltage decay first and then followed automatically for the negative. This mode is recommended for checking air ionizer equipment, as the Startbutton has to be pushed on time only to perform a complete check of ionizers including the determination of the Offset Voltage. 
     
    Offset-Voltage is the voltage potential, which remains after a certain time on the plate electrode. It is an indication for the balance of positive and negative ions. If the Offset- Voltage is positive, then an excess of positive ions emanate from the ionizer equipment - the other way round for negative. Please consider the specification of the ionizer for a qualified evaluation. 
     
    The Start-button has to be pressed in order to charge the plate-electrode by means of the integrated high-voltage unit. The timer setting and the applied voltage potential are shown on the display: 
     
    TIMER: 1000-100V 
    U = XXXXV 
     
    The timer starts as soon as the upper threshold value (1000 volt in this case) is passed: 
     
    TIMER=XX.Xs 
    U = XXXXV 
     
    and will stop the decay time when reaching the lower value (100 volt). The voltage setting for the timer could be changed in the set-up function. 
     
    The offset-voltage will be measured after a certain lapse time at the end of the static decay time measurement: 
    OFFSET=XXXs 
    U = XXXXV 
     
    The measurement series is now finished and all relevant parameters can be obtained from the display, e.g. to qualify an air ionizer equipment. The offset-voltage UOFF and the belonging lapse time will be finally displayed in the first row for 10 seconds or until a new measurement was initiated by pressing button A. 
     
     The decay time appears in the second row as POS DECAY or as NEG DECAY or both  alternating every second, if the auto-mode was pre-selected: 
     
    UOFF=XXXV T=XXXs 
    POS.DECAY =XX.Xs 
     
    2.3  Saving measurement values  
    The measured values can be saved in the internal memory, therefore the request will follow after each measurement accordingly:  
    SAVE VALUES? 
    A<YES NO>B 
     
    press the button A to start the saving. In case that the memory is full (99 entries), the display will indicate for approximay 2 seconds: 
     
    MEMORY FULL! 
     
    All files have to be deleted in order to perform the saving of the last measurement: The following request will appear on the screen:  
    DELETE FILES? 
    A<YES NO>B 
     
    In case, that the full memory should not be deleted, please press B and the unit will be ready to take the next measurement - the last measurement results will disappear and cannot be saved. 
     
    Important: It is recommended to transfer all files to your PC after each measurement interval by means of the enclosed software. Thus will allow to delete files of 
    the CPM374 at the next measurement interval without loosing important data. 
     
    Press A to delete the full memory of the CPM374 and the proposal for the file number of the last measurement will appear: 
     
    FILE NO: XX 
    A<OK CHANGE>B 
     
    Press A to confirm this file number or change with each press of B until you reach the right file number. Finally, press A to confirm the right file number. 
    In order to reach faster a certain file number press B, then the ten digit can be changed by pressing A while the single digit can be changed with B. when you have reached the right file number, wait for approximay 3 seconds until the display change back as stated above and confirm the setting by pressing A. 
     
    Press A and B simultaneously to start the set-up function. The CPM374 will ask for performing the set-up function:  
    SET-UP ? 
    A<YES NO>B  
     
    3.1  SET-UP  
    If the set-up function was started, the last Static Decay and Offset setting will appear (in this case Mode = Auto), which can be confirmed again by pressing A:  
    MODE = AUTO 
    A<OK CHANGE>B 
     
    If B will be pressed, the following options are shown one after the other: 
    MODE = POS 
    MODE = NEG 
    MODE = VOLTMETER 
    MODE = AUTO 
     
    The decay time for only positive respectively negative potentials with the following offset voltage are measured in the MODE = POS and MODE = NEG. In the MODE = 
    VOLTMETER, the CPM374 will measure permanently the voltage potential on the plate electrode - Use a measurement lead connected to the plate electrode to operate the 
    CPM374 as an Electrometer (High Impedance Voltmeter), e.g. monitoring the change of voltage potential on human bodies, caused by movements. 
     
    If you have selected the decay time measurement (MODE = POS, MODE = NEG or MODE = AUTO), the starting point for the voltage potential, from where on the decay time 
    should be measured, will be asked first:  
    START: U=1000V 
    A<OK CHANGE>B 
     
    Press A to confirm the proposed starting point. If B will be pressed, the following options are shown one after the other:  
    500V - 550V - ... - 900V - 950V - 1000V - 500V 
     
    and the selected value has to be confirmed by pressing A . 
     
     
    3.  SET-UP Function  
    The same procedure will follow selecting the stop point, that means at which potential the decay time should be stopped:  
    STOP: U=100V 
    A<OK CHANGE>B 
     
    Press A to confirm the proposed stop point. If B will be pressed, the following options are shown one after the other:  
    0V - 50V - 100V - ... -400V - 450V - 500V - 0V 
     
    and the selected value has to be confirmed by pressing A .  
    Finally, the CPM 374 will ask for the lapse time to determine the offset voltage:  
    OFFSET: XXXs 
    A<OK CHANGE>B 
      
    Press A to confirm the proposed lapse time. If B will be pressed, the following options are shown one after the other:  
    10s - 20s - ... - 50s - 1min - 2min- ... - 10min - 10s 
     
    and the selected value has to be confirmed by pressing A  . 
    The chosen parameters for Start, Stop and Offset will be shown for approximay 3 sec, 
    e.g.: 
    TIMER: 1000-100V 
    OFFSET: XXXs 
     
    and finally the unit will recheck if all parameters are chosen correctly:  
    TIMER+OFFSET OK? 
    A<YES NO>B 
     
    The selected parameters will be confirmed by pressing A and could be changed again with B. 
    When A was pressed, the unit will ask for clearing the memory, as the stored data files are based on the previous setting:  
    CLEAR MEMORY? 
    A<YES NO>B 
     
     
    If B was pressed, the previous setting will remain the same, the memory was not deleted and the new parameter change is ignored, the following message will appear: 
     PARAMETER 
    NOT CHANGED ! 
     
    If A was pressed, then following message will advice that the parameter change was taken over and the memory is cleared accordingly, the following message will appear for approximay 2 sec:  
    PARAMETER 
    CHANGED ! 
     
    The display will show now the new decay time setting and the request to press start! 
     
    3.2  Calibration  
    If NO>B was chosen for the set-up function, the unit will ask for the Electro-Fieldmeter calibration:  
    EFM CALIBRATION? 
    A<YES NO>B 
     
    Press B if a calibration is not required. Press A to initiate the automatic offset calibration, the following message will appear for approximay 1 sec:  
    PLEASE WAIT 
    CALIBRATION 
     
    Important: The plate electrode must be free of foreign voltage connection during calibration. 
     
    The display will show the new decay time setting and the request to press start! 
     
    If the electrofield-meter is defected, the following message will appear:  
    EFM DEFECT! 
    REPAIR IS NEEDED 
     
    If the power of the NiCd Battery is less than 6,7 volts, the unit will provide the following message alternating every second with the actual first row: 
     
    LOW BATTERY!  
     
    4.  Other messages on the display 
    The battery has to be recharged, but the ongoing or the upcoming measurement could be finished first.  
     
    Important:  
    Please use the enclosed battery charger only for charging the batteries. The batteries will also be charged when the CPM374 is switched off. The max. charging time is 14 hours, but the unit could be longer connected to the battery charger as it consists of a charging control device. It is recommended to use the CPM374 until LOW BATTERY! appears on screen. This will maintain the max. battery capacitance over a longer period. 
     
    If the battery voltage will become less than 6,1 volts, the following message will appear on screen for approximay 2 seconds:  
    AUTO OFF 
    LOW BATTERY! 
     
    and the unit will switch off automatically to prevent a total discharge of the battery. A total discharge can cause damage or destroy the battery! 
    Connect the unit by means of the provided cable to a COM socket of a PC. Insert the floppy disk and select the right drive (e.g. A:). Type ?START“ and press ENTER, the software will be installed on your PC. 
     
    Requirements: 
    min. 486 processor, 3.5“ floppy disk drive, 2 MB RAM, DOS or Windows. the program is self explaining. 
    We provide 12 month warranty from date of receiving the unit, if the unit was handled properly. The damage of the NiCd-battery caused by improper handling as well as 
    mechanical damage on the unit are excluded from warranty. Warranty is not provided, if the housing was opened. 
     
    5.  PC-Operating 
    6.  Warranty 
    The unit has to be grounded sufficiently for accurate measurements of the electrical field and its polarity. Therefore the unit has to be connected to ground potential by means of its grounding socket. 
     
    Housing :  Aluminum-housing 152mm x 152mm x 152mm (L x W x H) 
    Weight :  1,5 kg 
    Display :  Alphanumeric Display, 2 rows with 16 digits each, 
    display area 100 mm x 24 mm 
    Measurement Range :  Voltage potentials up to ± 1200 V, accuracy ± 2,5% 
    PC-Interface :  Serial port, 9 pol Sub D – socket 
    Battery :  7,2 V / 850 mAh 
    last approximay 4 hours for continuous operation when 
    battery was fully charged, charging time max. 14 hours 
    Battery Charger :  12 V / 500 mA 
    Load Current:  max. 350mA, max. 150 mA for battery operating 
    Operating Temperature:  0 ... 40°C 
    Relative Humidity:  0 ... 60% 
     
    ·  CHARGE PLATE Monitor CPM 374 
    ·  Power transformer prim: 230VAC sec: 12VDC 500mA 
    ·  Teflon insulated measurement lead, 1m 
    ·  Crocodile clip 
    ·  Grounding cable 
    ·  Serial PC-port cable 
    ·  3 Spacers for horizontal decay time measurement 
    ·  Operating Instructions 
    ·  Certificate of factory calibration 
     
      
    7.  Grounding 
    8.  Specification 
    9.  Parts of Delivery 
     
    The measurement unit is not approved for the use in explosive areas! 
    The unit is not approved for the use in power utilities! 
    This unit is not suitable to measure alternating fields with a frequency higher than 1 Hz. 
    Sparking discharge to the electro-fieldmill-sensor has to be avoided. 
     
    The Teflon-supports of the plate electrode must be maintained clean, therefore do not touch with fingers. All Teflon spacers and the teflonisolated wire to the plate have to be cleaned carefully with Acetone or similar solvents in case that the voltage on the charged plate drops in a rate of greater than 100V / minute under normal circumstances (no external lead connection to the plate electrode and no presence of air ionizers). The interval for factory calibration is normally 1 year. 
     
     
    Please note the following advice when checking air ionizers: 
    The presence of charged objects as well as personnel could cause failure on the measurement results caused by influence of voltage. The unit should be therefore always 
    grounded to avoid this phenomena from the housing of the unit itself. 
     
     
    10.  Safety Advice 
    11.  Maintenance 
    12.  Important Notice 
     
    KONFORMIT?TSERKL?RUNG 
     
    Die Firma 
     
    Kleinw?chter GmbH 
    Krummattstr. 9 
    79688 Hausen i.W. 
     
    erkl?rt in alleiniger Verantwortung, dass das Produkt 
     
    Charge Plate Monitor CPM 374 
    Messger?te zur Messung des elektrischen Gleichspannungsfeldes 
     
    auf die sich diese Erkl?rung bezieht, die Forderung 
     
    der EWG-Richtlinie über die elektromagnetische Vertr?glichkeit 
    Richtlinie des Rates vom 3.5.1989 (89/336) Stand: Juli 1993 
     
    erfüllt, insbesondere der Normen 
     
    EN 61010 ?Safety“ 
    EN 55011 Gruppe 1, Klasse B 
    EN 50082-2/92 
     
    für ISM-Ger?te. 
     
    Die Firma Kleinw?chter GmbH h?lt folgende technischen Dokumentationen zur Einsicht bereit 
     
    ·  vorschriftsm??ige Bedienungsanleitung 
    ·  Baupl?ne 
    ·  technische Dokumentationen 
     
    Hausen, den 09.01.2002 
     
     
    Jürgen Brunner 
    techn. Gesch?ftsführer 
     
     
    KLEINWACHTER
    Forschungs-. Entwlcklungs-. Produktlons-u. Vertrlebsges.m.b.H


    中國電子行業儀器儀表供應商——上海堅融實業有限公司      :www.adsenseban.net  
    凌志堅 產品     :        
    :jet1718  地址:上海市閔行區閔北工業園區紀王鎮紀西周涇 9 號/68 號 

    上一篇:德國Kleinwachter TOM-600靜電電阻測試儀手冊說明書

    下一篇:DESCO50671/50672離子風機操作說明書手冊

    產品分類

    聯系我們

    在線客服:

    主站蜘蛛池模板: 国产精品无码一区二区三级 | 青娱乐国产官网极品一区| 乱色精品无码一区二区国产盗| 国产一区二区内射最近更新| 国产免费av一区二区三区| 中文字幕一区二区精品区| 久久久人妻精品无码一区| 天海翼一区二区三区高清视频| 中文字幕永久一区二区三区在线观看| 性色av无码免费一区二区三区| 亚洲Av无码一区二区二三区 | 国产精品一区不卡| 亚洲欧美日韩国产精品一区| 末成年女AV片一区二区| 麻豆aⅴ精品无码一区二区| 精品一区二区三区色花堂| 日韩一区二区三区精品| 一区二区三区免费在线观看| 国产一区二区三区手机在线观看| 日韩AV无码一区二区三区不卡| 国模视频一区二区| 国产精品免费综合一区视频| 国产成人免费一区二区三区| 中文字幕AV一区二区三区 | 韩国精品福利一区二区三区 | 日本一区二区三区在线网| 国产精品一区二区香蕉| 精品国产一区二区三区香蕉| 国产日韩AV免费无码一区二区| 免费国产在线精品一区| 怡红院一区二区在线观看| 精品一区二区三区在线成人 | 国产拳头交一区二区| 无码一区二区三区在线| 多人伦精品一区二区三区视频| 精品不卡一区中文字幕| 精品人妻一区二区三区毛片 | 无码人妻一区二区三区精品视频 | 亚洲人成网站18禁止一区| 久久精品国产一区二区| 中文字幕日韩一区二区三区不|